+86-18822802390

Struzzjonijiet għall- użu ta ' PAL-1 Brix Meter

Nov 02, 2022

Instructions for use of PAL-1 Brix Meter


1. Aġġustament żero

1. Prepare distilled or tap water;

2. Naddaf il-wiċċ tal-priżma;

3. Drop about 0.3ml of water onto the prism surface;

4. Agħfas il-buttuna "START", il-vleġġa fuq il-wiri tteptep tliet darbiet u l-valur Brix se jintwera;

5. If it shows 0.0 percent , it means that the zero adjustment has been successful;

6. Jekk il-valur muri ma jkunx 0. 0 fil-mija , agħfas il-buttuna "ZERO" biex tirrisettja għal żero;

7. The instrument displays 0000, and the zero adjustment is completed. Wipe the water on the prism with a soft cloth or paper towel. Now the sample can be measured.

2. Test samples

1. Clean the prism surface;

2. Qatra madwar 0.3ml tal-kampjun ittestjat fuq il-wiċċ tal-priżma;

3. Agħfas il-buttuna "START";

4. Il-wirja turi l-valur Brix;

5. Il-valur Brix se jkompli jintwera għal minuta. Biex titfi, agħfas u żomm il-buttuna START għal aktar minn 2 sekondi;

6. Wipe off the sample, drop some water to the prism and dry it.

Nota: Tużax kuċċarina tal-metall meta żżid kampjuni mal-wiċċ tal-priżma biex tevita li tobrox il-wiċċ tal-priżma.

3. Messaġġ ta 'żball

AAA: Żball ta' żero.

1. During zero adjustment, no distilled water is placed on the prism surface.

2. When zeroing, other solutions are placed on the prism surface instead of distilled water.

LLL: Żball fil-kampjun.

1. Matul it-test, ma tpoġġa l-ebda kampjun.

2. Battery power is low.

Sample measurements above 53 percent (PAL-1)

Prism surface temperature is lower than 10 degree or higher than 40 degree


1. Sugar Content Measuring Instrument

Ibgħat l-inkjesta